The U.S. Food and Drug Administration is warning people to stop using certain types of glucose monitor sensors after the company that makes them, Abbott Diabetes Care, said the devices were linked to seven deaths and more than 700 injuries.
Certain FreeStyle Libre 3 and FreeStyle Libre 3 Plus sensors may provide incorrect low glucose readings, FDA officials said this week. Such readings over an extended period may lead people with diabetes to make bad treatment decisions, such as consuming too many carbohydrates or skipping or delaying doses of insulin.
“These decisions may pose serious health risks, including potential injury or death," the FDA said in the alert.
The sensors are devices that measure glucose levels in fluid just beneath the skin to provide real-time measurements of sugar in the blood. Information from the sensor is sent wirelessly to a device or phone.
The warning affects about three million sensors in the U.S. from a single production line, Abbott officials said in a statement. About half those devices have expired or been used, the company added. As of Nov. 14, the company reported seven deaths worldwide and 736 serious adverse events. No deaths occurred in the U.S., where 57 injuries were reported.
Abbott has notified all customers of the problem. The company said it has identified and resolved the issue in the affected production lot.
The FDA said people should stop using affected sensors and discard them.
The problem involved FreeStyle Libre 3 sensors with model numbers 72080-01 with unique device identifiers 00357599818005 and 00357599819002. It also involved FreeStyle Libre 3 Plus sensors with model numbers 78768-01 and 78769-01 and unique device identifiers 00357599844011 and 00357599843014.
People can visit www.FreeStyleCheck.com to check if their sensors are potentially affected and request a replacement, the company said. No other FreeStyle Libre products are affected.
——
The Associated Press Health and Science Department receives support from the Howard Hughes Medical Institute’s Department of Science Education and the Robert Wood Johnson Foundation. The AP is solely responsible for all content.
LATEST POSTS
- 1
This Canadian crater looks like marbled meat | Space photo of the day for Jan. 6, 2026 - 2
Europe: 4 Urban communities for a Paramount Social Experience - 3
The most effective method to Plan an Incineration Administration: A Bit by bit Guide. - 4
75% of Arab Israelis support Arab party joining government coalition post-war, survey reveals - 5
Italian court approves extradition to Germany of Ukrainian suspect in Nord Stream pipeline blast
Jersey's wellbeing score is below UK and France
Invigorating Spots To Go Kayaking All over The Planet
Sudan war ‘being fought on women’s bodies’: Survivors detail sexual assault
A24's 'Backrooms' trailer shows endless fluorescent-lit spaces and terrifying mannequins melting into the floor
The most effective method to Consummate the Specialty of Handshaking in Business and Group environments
What you need to know about Trump accounts as Michael and Susan Dell donate $6 billion to the new early childhood investment program
Don’t let food poisoning crash your Thanksgiving dinner
Aluminum salts emerge as likely target as health officials scrutinize childhood vaccines
How will the universe end?